EasyManuals Logo

Sartorius LE series User Manual

Sartorius LE series
84 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #22 background imageLoading...
Page #22 background image
22
External Calibration
Step Key (or instruction) Display
1. Tare the balance. w
0.0 g
2. Start calibration. q
+ 5000.0 g
CAL G
3. Apply the prompted calibration
weight (in this example, 5000 g).
4. The balance is calibrated 0.2 g
(displayed only if menu CAL ± G
code 1 10 2 is set).
5. If the “Calibrate, then
Adjust
*
auto adjust" setting is CAL
selected (1 10 1), the balance
is now adjusted automatically.
6. The calibration sequence is completed.
CC
CAL
7. After calibration/adjustment, the
+ 5000.0 g
weight is displayed with weight unit
8. Remove the calibration weight.
0.0 g
* Adjust shown only if menu code 1 10 2 is set.
Important note:
Afterwards, do not perform internal calibration/adjustment again.

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Sartorius LE series and is the answer not in the manual?

Sartorius LE series Specifications

General IconGeneral
BrandSartorius
ModelLE series
CategoryScales
LanguageEnglish

Related product manuals