22
External Calibration
Step Key (or instruction) Display
1. Tare the balance. w
0.0 g
2. Start calibration. q
+ 5000.0 g
CAL G
3. Apply the prompted calibration 
weight (in this example, 5000 g).
4. The balance is calibrated – 0.2 g
(displayed only if menu  CAL ± G
code 1 10 2 is set).
5. If the “Calibrate, then 
Adjust
*
auto adjust" setting is CAL
selected (1  10  1), the balance 
is now adjusted automatically.
6. The calibration sequence is completed.
CC
CAL
7. After calibration/adjustment, the 
+ 5000.0 g
weight is displayed with weight unit
8. Remove the calibration weight.
0.0 g
* Adjust shown only if menu code 1 10  2 is set.
Important note: 
Afterwards, do not perform internal calibration/adjustment again.