EasyManua.ls Logo

Sartorius LP Models - Practical Example: Animal Weighing; Determining Animal Weight with Automatic Start

Sartorius LP Models
99 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
44
Practical Example
Determining Animal Weight with Automatic Start of 20 Subweighing Operations for Averaging;
Automatic Printout of the Number of Subweighing Operations and of the Animal Weight
Step Key (or instruction) Display/Output
1. Select the animal weighing see “Preparation”
application in the Setup menu
2. Set the following animal weighing
parameters in the Setup menu:
– Animal activity: Active Setting 3 7 3 in Setup menu
– Start animal weighing: Automatic Setting 3 8 2 in Setup menu
– Printout: On: anim.wt.*fact. Setting 3 9 3 in Setup menu
– Print: All parameters Setting 7 1 2 in Setup menu
3. Prepare a container (cage) Place empty cage on
the balance
4. Tare the balance t
5. Enter number of subweighing 2 0
operations for averaging
6. Save number mdef soft key w
7. Weigh the first animal Place 1st animal in cage weight value fluctuates due to animal activity
8. Start automatic animal weighing start soft key v
mdef= x10 start
mdef= x10 start
mdef= 20 start
mdef= 20 start
M =-20xxxxx---
mul mdef S-id

Related product manuals