52
Factory Settings
Print application parameters
(automatic output of application
parameters): off (7 1 1)
Line format: for other apps/GLP
(7 2 2)
Preparation
Turn on the balance: Press e
> All display segments light up
Select the Recalculation applica-
tion in the Setup menu: Press M
Select menu:
Press the menu soft key (w)
Set parameter 2 1 9:
see “Configuring the Scale”
Exit the Setup menu: Press M
Setting Parameters for the
Recalculation Application
$ Access the Setup menu:
Select MENU
Set parameters for:
Automatic printout:
7 1 Print app. para.
Line format:
7 2 Line format
see “Configuring the Balance”
Exit the Setup menu: Press M
Additional Functions
In addition to the functions for:
alphanumeric input,
taring (not during alphanumeric
input), and
printing,
you can also access the following
functions from this application:
calibration
(not during alphanumeric input)
setup
turning off the balance.
Calibration/Adjustment
Press q
> See “Calibration/Adjustment”
for further instructions
Setup (setting parameters)
Press M
> See “Configuring the Balance”
for further instructions
Turning Off the Balance
Press e
> The balance shuts off
> The display goes blank