24
Drying Parameter Features
Heating temperature set to defined value
during measurement
Start of analysis
With stability after the
key is
pressed:
When START is shown in the function
line and you press
to confirm, the
initial weight is stored at stability
regardless of whether the cover is open
or closed.
Measurement begins as soon as the cover
is closed.
Without stability after the cover is
closed:
A symbol shown in the graphic symbol
display prompts you to close the cover
once the initial weight condition is met.
The initial weight is stored without
stability as soon as the sample chamber
is closed, and analysis begins.
End of Analysis with Shutoff
Parameters
Fully automatic mode
Timer mode
Fully automatic mode:
Use the fully automatic mode when
loss of weight on drying follows a
clearly delineated curve which can be
unambiguously evaluated (see below).
Weight
Moisture (evaporation)
Automatic shutoff
Time
Timer mode:
The analysis ends as soon as the
specified time has elapsed.
Display Mode for Result
The following units can be selected for
displaying analysis results:
– Moisture %M
Dry weight %S
– Ratio %MS
Residual weight g
Print Intermediate Results
Intermediate results can be printed
either at user-definable intervals or by
pressing the r key.