38
Overview
Specifications
Weighing capacity (Max) 35 g
Typical reproducibility of the
weighing system
Starting at an initial sample weight of approx. 1 g: ± 0.2%
Starting at an initial sample weight of approx. 5 g: ± 0.05%
Readability 1 mg; 0.01%
Typical sample quantity 5 – 15 g
Display of results % moisture
% dry weight
% ratio
g residual weight
Shutoff parameter optional:
– Fully automatic
– Manual shutoff
– Timer settings 0.1–99 min
Sample heating Even sample heating capability with an IR dark radiator
(metal tube heater)
Access to sample chamber – through hinged, flip-up cover with wide opening angle
– with viewing window
Draft shield Integrated draft shield
For conformity with FDA/HACCP regulations Aluminum panels (in place of glass panels)
Operating temperature range and setting 40°C to 160°C (104°F to 320°F), adjustable in 1°C increments
Operator guidance Symbols representing corresponding functions
Heating program Standard drying
Measuring program 1 program saved in a non-volatile memory (freely selectable
parameters)
Memory for data storage Results are saved until the start of the next measurement
Log printout – Inalterable standard configuration GLP report for
measurement results/calibration/adjustment
– Short report for measurement results minimizing
paper usage
– Using an optional external printer
Data interface RS232 C-S/VA24-V28, 7 data bits (ASCII) 1 parity bit
Monitoring and control of the accuracy of
the analyzer as inspection, measuring and
test equipment
External calibration with optional calibration weights;
Temperature adjustment with an optional adjustment set
Housing dimensions (W × D × H) 224 × 366 × 191 mm
Net weight approx. 5.8 kg
Power requirements (supply voltage) 230 V or 115 V (depending on the model) (–15% … +10%)
Frequency 50 – 60 Hz
Fuses 2 (zero conductor/phase), 6.3 A, time-lag (slow-blow),
5 + 20 mm (internal)
Power consumption max. 400 VA