Operating Instructions PMA 35001-X 15
Applications
3. Add first component.
+ 50.0 g
2. Press the U key
(Zero/Tare).
0.0 g
Example (Cumulative Weighing)
1. Center an empty paint can
on the load plate.
+ 118.0 g
4. Press the L key [MEM]
STO 01
5. Add the second component
+ 110.0 g
6. Press the L key [MEM]
STO 02
7. Add the third component
+ 203.0 g
Oops! You poured in too
much. The correct value for
the formula is 200.0 g.
8. PPress the S key to start
the recalculation mode.
A »C« (“Correction”) flashes
on the display.
9. Press the S key repeatedly
to correct the value.
+ 200.0 g
10. Press the L key [MEM]
COR 01
11. Add the required amount of
the first component. “C1”
appears on the display.
–1.7 g.
12. Add until value reaches
0.0 g.
0.0 g
13. Press the L key [MEM]
COR 02
14. Add the required amount
of the second component.
“C2” appears on the display.
2.0 g
15. Add until value reaches
0.0 g.
0.0 g
16. Press the L key [MEM].
The scale returns to the
formulation program. »C«
is no longer displayed.
+ 200.0 g.
17. Press the c key (7) [REC]
to view the factor by which
the total weight will exceed
the original target.
(C = “Correction;”
in this example, 1.03).
(Total weight = original
target + correction factor)
18. Add fourth component
+1000.0 g
This concludes the
recalculation example.