SARTORIUS Portable and Gold
29
- Insert the membrane-covered keypad (120.A) into plug-in position Bu.8, the
strain gauge system connecting cable (106.B) into position St.1 of the main
circuit board (118).
- Now perform a complete adjustment of the balance/scale:
Adjust the overload stops (see page 16)
Adjust the zero point offset value (see page 17)
Check the off-center load (see page 18)
Adjust the linearity (see page 19)
Calibrate the balance/scale (see page 23)
- After completing the adjustment procedures, close the balance/scale housing.
- The balance/scale is now ready to operate.