8
Table of Adjustment Data
Balance/scale
type
PT 120
-000V1
PT 120
-000V103
PT 210
-000V1
PT 600
-000V1
PT 1200
-000V1
PT 2100
-000V1
PT 6
-000V1
GT 210
-G00V1
GT 600
-G00V1
Weighing
capacity
121 g 210 g 610 g 1210 g 2100 g 6100 g 210 g 610 g
Readability
0,01 g 0,1 g 1 g 0,01 g 0,1 g
Reproducibility
¹·¹ )
<=± 0,02 g <=± 0,1 g <=± 0,2 g <=± 2 g <=± 0,02 g <=± 0,1
Linearity
<=± 0,01 g <=± 0,1 g <=± 1 g <=± 0,01 g
Calibration
weight
100 g / M1 500 g / M1 1000 g / M1 5000 g / M1 100 g / M1 500 g /
Off-center test
weight
100 g
ø = 28 mm
200 g
ø = 35 mm
500 g
ø = 47 mm
1000 g
ø = 50 mm
2000 g
ø = 65 mm
5000 g
ø = 80 mm
200 g
ø = 35 mm
500 g
ø = 47 mm
ø
Off-center test
area
up to the edge of the
weighing pan
Off-center
tolerance ¹·² )
<=± 0,04 g <=± 0,05 g <=± 0,3 g <=± 3 g <=± 0,05 g <=± 0,3 g <
Zero point offset
test weight
balance/scale unloaded
Zero point offset
control
8440 8080 10180 8440 8080
Zero point offset
tolerance
5450 ... 11440
3890 ...
12275
7780 ...
12275
5450 ...
11440
3890 ..
12275
¹·¹) Reproducibility in conformance with OIML; balance
loaded five times with a weight near the maximal
load.
¹·²) The off-center tolerances are production data. As a
result of transport, differences in installation conditions
and measuring methods, deviations of two digits
beyond the specified tolerances are admissible for
standard instruments in the as-delivered condition.