Specifications
User Manual Secura
®
, Quintix
®
, Practum
®
15. Specifications
15.1 General Data
AC adapter
Sartorius AC Adapter YEPS01-PS4 or YEPS01-PS5 with replaceable country-specific power plug adapters
Primary 100 240 V~, ± 10 %. 50 60 Hz ± 5 %, 0.2 A
Secondary 15 V
DC
, ± 5%, 530 mA (max.) / 8 Watt (max.): 0°C to +40°C
15 V
DC
, ± 5%, 330 mA (max.) / 5 Watt (max.): +40°C to +50°C
Further details Protection class II in accordance with EN/IEC 60950-1 | up to 3000 m above sea level
IP40 in accordance with EN 60529/IEC 60529
Balance
Supply voltage Only via Sartorius AC adapter YEPS01-PS4 or YEPS01-PS5 with replaceable country-
specific power plug adapters
Power consumption W 4.5 (typically)
Further details IP43 in accordance with EN 60529/IEC 60529
Ambient conditions
The technical data apply under the following ambient conditions:
Environment For indoor use only
Ambient temperature* °C +10 to +30
Operational capability °C Guaranteed between +5 and +45.
Storage and shipping °C –10 to +60
Height above sea level (NN) m Up to 3000
Relative humidity** 15 80% for temperatures up to 31°C non-condensing, decreasing linearly up to 50%
relative humidity at 40°C and 20% at 50°C.
Safety of Electrical Equipment In accordance with EN 61010-1/IEC 61010-1 Safety requirements for electrical equipment
for measurement, control and laboratory use – Part 1: General requirements
Electromagnetic Compatibility In accordance with EN 61326-1/IEC 61326-1
Electrical equipment for measurement, control and laboratory use – EMC requirements –
Part 1: General requirements
Interference resistance: Suitable for use in industrial areas
Transient emissions: Class B (suitable for use in residential areas and areas that are connected to a low voltage
network that (also) supplies residential buildings). The device can therefore be used in both
areas.
Standard Equipment
Selectable weight units
1
Gram, kilogram, carat, pound, ounce, troy ounce, Hong Kong tael, Singapore tael,
Taiwanese tael, grain, pennyweight, milligram, parts per pound, Chinese tael, momme,
Austrian carat, tola, baht, mesghal and Newton
Available application programs Secura
®
models: Counting, percentage, density determination, peak hold, unstable
condition, checkweighing, mixing, components (totalizing), statistics, conversion, SQmin,
and identifiers
Quintix
®
models: Counting, percentage, density determination, peak hold, unstable
condition, checkweighing, mixing, components (totalizing), statistics, conversion
Practum
®
models: Counting, percentage, density determination, peak hold, unstable
condition, checkweighing
Verified balances in accordance with EU requirements comply with the requirements of Council Directive 2014/31/EU with
EN 45501:2015 and OIML R76:2006.
* For verified balances in accordance with EU requirements, refer to the information on the balance.
** For verified balances in accordance with EU requirements, the legal regulations apply.
1) Depending on the country-specific model version, not all weight units listed may be available.

Table of Contents

Other manuals for Sartorius Quintix

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Sartorius Quintix and is the answer not in the manual?

Sartorius Quintix Specifications

General IconGeneral
DisplayBacklit LCD
Response Time1.5 s
Operating Temperature+10°C to +30°C
Readability0.1 mg to 1 g
CalibrationInternal/External
ConnectivityRS-232, USB
Repeatability0.1 mg
Linearity±0.2 mg
Operating Humidity15% to 80% non-condensing
Power Supply100 – 240 V, 50/60 Hz
Pan SizeVaries by model, typically 90 mm diameter (round) or 190 x 204 mm (rectangular)