Safety advice for servicing Sartorius Competence balances/scales.
Outlines key service procedures like repeatability, off-center error, and linearity checks.
Explains how to adjust the zero-point offset value using a DVM and balancing resistor R29.
Details checking repeatability for double lever systems by calculating standard deviation from measurements.
Describes how to check for off-center loading errors by placing test weights in different positions.
Explains checking off-center errors for CP and TE balances with different weighing pans.
Explains the procedure for adjusting off-center errors by filing a thin part on the strain-gauge load receptor.
Details using service software to adjust linearity after span adjustment.
Guides on adjusting off-center loading error using a 3-point measurement and adjustment screws on double-lever systems.