Sartorius Talent TE12000 Specifications

General IconGeneral
BrandSartorius
ModelTalent TE12000
CategoryScales
LanguageEnglish

Summary

Generals Information

Important Information

Crucial safety and qualification requirements for servicing Sartorius balances.

General Notes

Activating the BPI Mode

Procedure for Connecting the Adapter

Step-by-step guide to connect the BPI adapter for service software use.

Balances/Scales with Strain Gauge Systems

Checking the Overload Stop

Procedure for checking and resetting the overload stop using service software.

Zero-point Offset Value

Adjusting the Zero-point Offset Value

Steps to adjust the zero-point offset using a digital voltmeter and balancing resistor.

Repeatability (Standard Deviation)

Off-center Error

Procedure for checking and adjusting the off-center loading error.

Span Adjustment

Adjusting the Span

Instructions for span adjustment using service software or calibration weights.

Adjusting Linearity

Procedure for checking and adjusting the linearity of the balance.

Balances/Scales with Double Lever Systems

Checking the Repeatability (Standard deviation ā€žSā€œ)

Method for calculating and checking repeatability for double lever systems.

Off-center Error

Adjusting the Off-Center Error

Procedure for Adjusting the Off-Center Loading Error (Double-Lever System with Adjustment Screws (P))

Steps to adjust off-center loading error using a 3-point measurement and adjustment screws.

Procedure for Adjusting the Off-Center Error (Double Lever System)

Filing at Thin Places

Instructions on adjusting off-center error by filing thin places on the upper guide.

Opening and Closing the Balance/Scale

Replacing the PCB

Replacing the Data Output Port

Error Codes

Display Codes and Solutions

Lists error codes, their causes, and suggested solutions for common issues.

Further Error Codes and Solutions

Continuation of error codes and their corresponding causes and solutions.