EasyManua.ls Logo

Sartorius Talent TE12000 - Span Adjustment; Adjusting the Span

Sartorius Talent TE12000
56 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
46
Span Adjustment
Note: There are two methods for adjusting the span:
1. Using the service software
2. As described in the following:
Select menu item 1 5 1, „Calibration/adjustment: Accessible“ in the
operating menu (factory setting). Make sure the menu access switch is
open
(see page 19).
If necessary, select the calibration weight unit in the operating menu (item 1
4 x; factory setting: 1 4 1*, „g“).
Unload the balance/scale (in this example, model TE124S) and press t to
zero it, if necessary.
Press and hold the t key (> 3 sec) to initiate the calibration routine.
The required calibration weight is displayed without a weight unit; e.g.,
+100.0000+100.0000
+100.0000+100.0000
+100.0000
Center the displayed weight* on the weighing pan/load plate
(in this example, 100 g)
Important: The balance/scale will only accept a weight that is within a tolerance range
of approximately 2% of the nominal weight. Greater deviations can be
corrected only using the service software.
After calibration, the readout shows the calibration weight* with weight
unit (in this example, „+ 100.0000 g“)
Remove the calibration weight.
Check the span and perform adjustment again if necessary.
* depends on model
Anz-151F.eps
Anz0000F.eps
Anz-100F.eps
Anz-100gF.eps