Capacity | 4100 g |
---|---|
Readability | 0.1 g |
Display | LCD |
Linearity | ± 0.2 g |
Calibration | External |
Power Supply | AC adapter |
Repeatability (Std. Dev.) | 0.1 g |
Stabilization Time (typical) | 2 seconds |
Operating Temperature | +10°C to +30°C |
Interface | RS-232 |
Instructions for servicing Sartorius Competence balances/scales, requiring prior service course completion.
Step-by-step guide for connecting the BPI adapter and activating BPI mode for service software use.
List of essential checks and adjustments: overload stop, zero point, repeatability, off-center error, span, linearity.
Procedure for adjusting zero-point offset using a digital voltmeter and balancing resistor R29.
Procedure for adjusting off-center error by filing a thin part on the strain-gauge load receptor.
Procedure for span adjustment using menu item 1 5 1 and a calibration weight.
Using service software to adjust linearity, requiring BPI mode and PC connection.
Essential checks: overload stop, repeatability, off-center error, span, and external linearity.
Procedure for adjusting off-center error using adjustment screws (P) on double-lever systems.
Guidance on filing thin parts to adjust off-center error, focusing on the largest error first.
Procedure for span adjustment using menu item 1 5 1 and a calibration weight for double lever systems.
Using service software to adjust linearity for double lever systems, requiring BPI mode.
Step-by-step guide for replacing the main PCB, including data backup and post-replacement checks.
Troubleshooting guide for error codes related to load capacity, pan placement, zeroing, and tare memory.