WTE5001-e03101
Service Manual
Sartorius Talent | Gem, Gold
TE Models
Electronic Analytical and Precision
Balances and Precious Metal Scales
Including Service Specification Sheets
Question and Answer IconNeed help?

Do you have a question about the Sartorius Talent TE4101 and is the answer not in the manual?

Sartorius Talent TE4101 Specifications

General IconGeneral
Maximum Capacity4100 g
Readability0.1 g
CalibrationExternal
DisplayLCD
Repeatability0.1 g
Linearity±0.2 g
InterfaceRS-232
Operating Temperature10 °C to 40 °C
Weight2.5 kg
Power SupplyAC adapter
Stabilization time2 seconds

Summary

General Information

Important Information

Crucial safety and authorization requirements for servicing Sartorius balances.

Service Concept for TE- Balances/Scales

Warranty and Defect Handling Procedures

Outlines warranty claims, procedures for defective units, and return protocols.

Maintenance and Adjustment Procedures

PCB and Housing Management

Procedures for exchanging the main PCB and opening the balance housing.

Activating the BPI Mode

Connecting the BPI Adapter

Step-by-step guide for connecting the BPI adapter to the balance/scale.

Completing BPI Mode Activation

Final steps to activate BPI mode and caution for exiting BPI mode.

Zero-point Offset Value Adjustment

Adjusting the Zero-point Offset Value

Describes the process of adjusting the zero-point offset using a DVM.

Table of Adjustment Resistors

Correlates zero-point offset voltage with resistor values for adjustment.

Span Adjustment

Adjusting Linearity

Adjusting Off-Center Error (Double-Lever System)

Procedure with Adjustment Screws

Details adjusting off-center errors using screws on double-lever systems.

3-point Measurement for Adjustment

Method for determining and adjusting off-center loading errors using 3-point measurements.

Adjusting Off-Center Error by Filing

Guidance on adjusting off-center errors by filing thin parts on specific models.

System Positioning for Adjustment

Off-center Loading Error Filing Positions

Illustrates filing positions for off-center error adjustment on different pan types.

Span Adjustment (Double Lever System)

Span Adjustment Procedure

Details span adjustment using menu settings and calibration weights for double lever systems.

Adjusting Linearity (Double Lever System)

Checking and Adjusting Linearity

Procedure for checking and adjusting linearity for double lever systems.

Replacing the PCB

PCB Replacement and Data Reprogramming

Procedure for replacing the main PCB and reprogramming data records.

Replacing the Data Output Port

Data Output Port Replacement and Testing

Steps for replacing the data output port and testing its functionality.

Error Codes and Solutions

Error Codes H, L, E01, E02, E10, E11

Lists error codes, their causes, and corresponding solutions.

Error Codes E22, E30, E50, E53, E54, E55

Lists additional error codes, their causes, and corresponding solutions.