Summary

Generals Information

Important Safety and Service Prerequisites

Mandatory service course completion for personnel servicing Sartorius Competence balances/scales.

Service Concept for TE- Balances/Scales

General Operating and Service Notes

BPI Mode and Write-Protect Control

Instructions for activating BPI mode and deactivating write-protection.

General Service Procedures

Write-Protect Reset and PCB Replacement

Procedure for resetting write-protect and guidelines for replacing the main PCB.

Off-Center Load and Span Adjustment

Methods for adjusting off-center load and span on balances/scales.

Menu Access Switch Operation

Balance/Scale Operating Menu

Modifying Menu Code Settings

Advanced Menu Code Settings

Menu 9: Reset Menu to Factory Settings

Procedure for restoring factory settings and resetting the SBI mode.

Activating BPI Mode for Service

BPI Adapter Connection and Setup

Step-by-step instructions for connecting the BPI adapter to the balance/scale.

Finalizing BPI Mode Activation

Equipment Connection and BPI Mode Activation

Connecting equipment and completing the BPI mode activation sequence.

Important: Write-Protect and SBI Mode

Warning about setting write-protect and returning to SBI mode after BPI use.

Detailed Overload Stop Check

Zero-point Offset Value Adjustment

Importance of Zero-point Offset Adjustment

Reasons why zero-point offset adjustment is necessary for strain-gauge load receptors.

Zero-point Offset Adjustment Procedure

Procedure for adjusting zero-point offset using a digital voltmeter and specific connections.

Adjustment Resistor Reference Table

Repeatability (Standard Deviation) Testing

Off-center Error Testing

Off-center Error Adjustment

Method for adjusting off-center error by filing a thin part on the load receptor.

Span Adjustment Procedure

Span Adjustment Methods Overview

Two methods for span adjustment: using service software or manual calibration.

Manual Span Adjustment Steps

Steps for manual span adjustment using menu items and calibration weights.

Linearity Adjustment Procedure

Double Lever System Performance Checks

Off-center Error in Double Lever Systems

Off-center Error Check for Pan Types

Adjusting Off-center Error in Double Lever Systems

Off-center Error Adjustment Procedure (Double-Lever)

Steps for adjusting off-center error using a 3-point measurement and adjustment screws.

Safety Warning: Open Balance/Scale Operation

Safety warning regarding working on an opened balance/scale with voltage present.

3-Point Measurement and Adjustment

3-Point Measurement for Adjustment

Detailed steps for performing a 3-point measurement to determine off-center loading error.

Correcting Deviations with Adjustment Screws

How to adjust screws to correct positive or negative deviations in off-center error.

Fine-Tuning Off-center Error (Double Lever)

Filing for Off-center Error Adjustment

Technique for adjusting off-center error by filing thin places on the upper guide.

Cautionary Notes for Filing Adjustments

Caution regarding filing adjustments and checking largest errors first.

Span Adjustment (Manual Method)

Linearity Adjustment (Manual Method)

Linearity Adjustment with Service Software

Using service software to adjust linearity after span adjustment.

Balance/Scale Opening and Closing Procedures

Main PCB Replacement Procedure

Main PCB Data Backup

Procedure for backing up data using service software before replacing the main PCB.

PCB Replacement and Reprogramming Steps

Steps for replacing the main PCB and reprogramming necessary values.

Data Output Port Replacement

Troubleshooting: Error Codes

Further Troubleshooting: Error Codes