46
Span Adjustment
Note: There are two methods for adjusting the span:
1. Using the service software
2. As described in the following:
– Select menu item 1  5  1, „Calibration/adjustment:  Accessible“ in the
operating menu (factory setting).  Make sure the menu access switch is
open
  
  
 (see page 19).
– If necessary, select the calibration weight unit in the operating menu (item 1
4  x; factory setting: 1  4  1*, „g“).
– Unload the balance/scale (in this example, model TE124S) and press t to
zero it, if necessary.
– Press and hold the t key (> 3 sec) to initiate the calibration routine.
– The required calibration weight is displayed without a weight unit; e.g.,
„
+100.0000+100.0000
+100.0000+100.0000
+100.0000“
– Center the displayed weight* on the weighing pan/load plate
(in this example, 100 g)
Important: The balance/scale will only accept a weight that is within a tolerance range
of approximately 2% of the nominal weight.  Greater deviations can be
corrected only using the service software.
– After calibration, the readout shows the calibration weight* with weight
unit (in this example, „+ 100.0000 g“)
– Remove the calibration weight.
– Check the span and perform adjustment again if necessary.
* depends on model