Checking and Adjustment Sequence
Checking the Overload Stop
Note: Overload stops must be checked only following replacement of the strain-
gauge system.
The overload stop must be reset after the strain-gauge load receptor has been
damaged or replaced. Weight readouts greater than the balance/scale’s
maximum capacity must be read off. The service software is required for this
procedure.
Activate the BPI mode (see page 26)
Service software: Select the „Diagnostics“ program; then select the „High
Resolution“ subprogram
When adjusting the overload stop, read off the weight value shown on the
Psion server or PC display (see page 30)
After adjusting the overload stop, activate the „Close“ function in the service
software to reset the balance/scale to the SBI mode.