EasyManua.ls Logo

Sartorius Talent TE12000 - Adjusting the Off-Center Error; Procedure for Adjusting the Off-Center Loading Error (Double-Lever System with Adjustment Screws (P))

Sartorius Talent TE12000
56 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
42
Adjusting the Off-Center Error
Note: Following adjustment of the off-center loading error, the span and linearity
must be tested and, if necessary, adjusted.
Procedure for Adjusting the Off-Center Loading Error (Double-Lever System
with Adjustment Screws (P))
Use a 3-point measurement to determine the off-center loading error.
The off-center loading error of the balance is indicated by the results of the
3-point measurement.
Note: The balance/scale must be opened to adjust the off-center loading error.
Open the balance/scale (see page 48).
Re-connect power, if disconnected.
Caution: From this point, you will be working on an opened balance/scale
while the equipment is carrying voltage.
The adjustment screws (see arrows) are now accessible.
Adjust both screws to set the off-center loading error
(see illustrations on the left).
To test the adjustment of the off-center loading error, close the balance/
scale and position the weighing pan/load plate on the balance/scale.
Allow approximately 3 to 5 minutes for the balance/scale to warm up.
Perform the 3-point adjustment as described on page 43.
Aut13353b.jpg
Aut13353a.jpg
PP
PP
P