37
Adjusting Linearity
Checking the Linearity
Note: The linearity must be checked according to the Standard Operating Procedure
WKD-038.
- Check the linearity of the balance/scale for the entire weighing range in 4-g
steps.
- Compare the values displayed with the tolerance ranges given in the „Table
of Calibration/Adjustment Data“ on pages 53-54.
- If the erors exceed the permissible tolerance ranges, the linearity must be
adjusted
Adjusting the Linearity
Use the service software to adjust linearity.
Activate the BPI mode (see page 26).
Connect the balance/scale to the PC and call the linearity program in the
service software.
Test the span and adjust if necessary (see page 36, „Span Adjustment“).
After linearity and span has been adjusted, activate the „Close“ function in
the service software to reset the balance/scale to the SBI mode.