14
If you have deactivated the automatic taring
function (-a2-):
§ Remove the sample from the weighing pan
§ Place the next sample on the weighing pan
§ Transfer the value by pressing the key
> The values will be printed out (see example)
If you have activated the automatic taring function (-a2-):
§ Leave the sample on the weighing pan
§ Add the next sample to the one already on the weighing
pan
§ Make sure that the expected total weight does not exceed
the maximum capacity of your weighing instrument
(otherwise, “H” will be displayed)
§ Transfer the value by pressing the key
> The values will be printed out
§ To record consecutive weighing operations, follow the
steps mentioned above. Then:
§ Press the key to obtain a printout of the statistics
> The values, as shown on the left, are automatically
computed and printed out
> Once the data has been printed out, the statistics memory
will be cleared
Vartext1, Vartext2
You can define a header consisting of 2 lines maximum,
each with 24 characters maximum. These user-defined
lines will be printed at the beginning of each hard copy,
and are identified as Vartext1 and Vartext2. They are
stored in the printer’s non-volatile memory (to input head-
ers, see “Settings” on page 27).
123456789012345678901234
001: + 123.6 g
002: + 117.8 g
003: + 119.1 g
123456789012345678901234
n 12
x 99.8 g
s 3.55 g
srel 3.57 %
x 1203.4 g
min 96.0 g
max 103.2 g
Diff 7.2 g
12-DEC-95 16:02:11