SciAps X-550 Pb QuickStart Guide
14
7.4 Classification Level
This level controls whether the analyzer
determines a Positive or Negative classification
on the screen. Tap to change.
Default is the Federal Pb level of 1.0. User may
change this based on local Pb limits.
Note: This is NOT the Action Level setting that
controls use of Quick Mode. The PCS specifies
that the action level must be set to 1.0
regardless of classification level. In the Lead
Paint application, the action level is fixed at 1.0
and user cannot change it.
7.5 Test Mode
OSHA/IH (Timed) – Test will run until set time
has elapsed (default 10sec).
PCS (Quick) – Test will run until analyzer can
make a clear Positive or Negative
determination, up to a maximum of the set
test time.
Note: This setting may affect what Pb
screening threshold applies or whether an
inconclusive range applies. Please consult PCS
for more information.
Action Level on this screen is shown for
informational purposes. User cannot change
this as all X550Pb PCS require this to be set to
1.0.