SciAps X-550 Pb QuickStart Guide
15
7.6 PCS Cal Check
Run additional check of 1.0 standard. The
Software will enforce this every 8 hours or
every time the analyzer is powered on. Users
may prefer to run the check more frequently.
7.7 Energy Calibration
Run additional clip calibration. This serves as a
system check. Software will require this to run
at startup. Users do not have to run this more
frequently.