SciCan HYDRIM
®
C61W G4 Quick Reference Guide
Operator’s Manual should be consulted before using this washer.
Load Instruments and close the door.
Press START.Select cycle.
Power on the unit.
Spare Parts and Accessories
CS-HIPC-U HIP- Ultra Cleaning Solution (8 x 750 mL)
01-112594S Water Softener Salt
01-108305S Water Test Kits
01-114390S HEPA Filters
WTI-SK Wash Test Indicator - Starter Kit
WTI-RK Wash Test Indicator - Refill Kit
CF2 Chamber filling failure
Check that the water hoses are not kinked.
Check that water shut-off valves are opened.
Call for Service.
CF23 Top wash arm low RPM
Top wash arm speed too low.
Check for obstructions to rotation of top wash arm.
Check that holes in wash arms are not plugged.
CF24 Bottom wash arm low RPM
Bottom wash arm speed too low.
Check for obstructions to rotation of bottom wash arm.
Check that holes in wash arms are not plugged.
BASIC OPERATION
Refilling the Water Softener Salt Reservoir
TROUBLESHOOTING
3
4
www.scican.com
Terminal sterilization of instruments MUST
follow processing in the HYDRIM C61W G4.
2
1. Unscrew lid.
2. Fill to the top (max. 0.5 kg / 1.1 lbs.)
3. Screw lid on tightly.
4. Run a complete cycle with no instruments.
IMPORTANT!
If the salt container is being filled for the first time, it MUST be filled
with at least 1 litre of water before adding the salt.
Only tablet salt should be used in the softening unit. Fine grained
salt such as table salt should NOT be used as it could clog the unit.
The water softener salt is also available from SciCan.
Please contact your local dealer.
Refer to Operator’s Manual for a more comprehensive list. Call for service if suggested recommendations do not resolve the error message.
1
1-800-667-7733
1-800-572-1211
Rinse and Hold Cycle
Prevents soil from drying on instruments.
Incomplete loads that will not be washed
within the hour.
NOT a cleaning cycle.
Regular Cycle
Moderately soiled loose instruments.
Heavy Duty Cycle
Heavily soiled instruments and cassettes.
TOUCHSCREEN
Cleaning the Chamber
P1
Rinse&Hold
P2
Regular
P3
Heavy Duty
USB
Door
Lock
User/Technical
Menu
Network
Connection
Water
Softener/
Detergent
Level/Door
Status
Enter
Back
Pour 0.5L of vinegar or citric acid-based instrument neutralizer into
the chamber and follow these steps:
P0 Cleaning
IMPORTANT! Do a Chamber Cleaning cycle every 25 cycles.
Failure to clean the chamber can promote corrosion.
MSDS available for download at www.scican.com
HYDRIM is a registered trademark and HIP is a trademark of SciCan Ltd. Manufactured by SciCan Ltd. 1440 Don Mills Road, Toronto, Ontario, Canada, M3B 3P9.
The button is inside the kick plate.
95-115868 NA EML R2
Replacing the HIP
Cleaning Solution
Recommended precautionary measures with regard to safe handling of HIP solution:
Wear undamaged surgical gloves and eye protection.
After handling the product, remove and dispose of the gloves and wash hands prior to removing eye protection.

Other manuals for SciCan HYDRIM C61w G4

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the SciCan HYDRIM C61w G4 and is the answer not in the manual?

SciCan HYDRIM C61w G4 Specifications

General IconGeneral
BrandSciCan
ModelHYDRIM C61w G4
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals