STATIM
STATIM
CASSETTE AUTOCLAVE
TROUBLESHOOTING GUIDE
Version 2.1
Original Statim Error Codes
The Statim Sterilizer should never leak steam or have water dripping from under
the machine. If either of these conditions occur, check your cassette and seal. All
leaks should be corrected before proceeding with troubleshooting procedures.
Check Cassette 1
Chamber temperature does not rise more than 2°C in three minutes. In all cases: No power to the
boiler. Check thermal fuse. Replace if necessary.
Check Cassette 2
http://www.sterilizers.com/statim-trouble-shooting-manual.html (1 of 26) [3/27/2004 9:23:44 PM]
Question and Answer IconNeed help?

Do you have a question about the SciCan STATIM and is the answer not in the manual?

Summary

Troubleshooting Guide

Original Statim Error Codes

Discusses original error codes for Statim sterilizers and initial leak checks.

Check Cassette 1

Addresses failure to rise in chamber temperature, indicating no power to the boiler.

Check Cassette 2

Deals with steam leaking from cassette during operation or seal issues.

Check Cassette 3

Addresses failure to pressurize to 110°C and potential steam leaks.

Check Cassette 4

Covers failure to begin sterilization within 10 minutes of pressurization.

Check Cassette 5

Relates to the boiler running dry prematurely, suggesting pump or boiler issues.

Check Cassette 6

Addresses boiler temperature exceeding chamber temperature during sterilization.

Check Cassette 7

Covers boiler temperature below lower limit during sterilizing.

Check Cassette 8

Deals with chamber temperature being higher than boiler during sterilizing.

Service Needed

Explains "Service Needed" message due to high temperatures or broken thermocouple.

New Statim Error Codes

Water Quality Issues

Covers "Water Quality is Not Acceptable" and "Distilled H20 Only" messages.

Maintenance and Repair Procedures

Changing the Cassette Seal

Instructions on how to replace the cassette seal for optimal performance.

Thermal Fuse Replacement

Procedure for removing and replacing the thermal fuse component.

Microswitch Replacement

Detailed steps for removing and installing the microswitch component.

Reclaiming Statim Boiler

Procedure for cleaning and restoring the Statim boiler, with a caution for sensors.

Boiler Calibration

Instructions for calibrating the steam generator (boiler) for accurate readings.

System Settings and Specifications

Diagnosing Water Quality Sensor Problems

Procedures for identifying and resolving issues with the water quality sensor.

SciCan STATIM Specifications

General IconGeneral
ManufacturerSciCan
Sterilization MethodSteam
Cycle Time Unwrapped6-9 minutes
Cycle Time Wrapped10-15 minutes
Water TypeDistilled water
ModelSTATIM
Sterilization Temperature134°C (273°F)