All attached regular and contra-angle handpieces and turbines are cleaned inside in turn. After a successful process, the
program slots will be displayed inversely. The display shows: "Cleaning successful".
Open the front door.
Remove regular and contra-angle handpieces by pressing the locking button.
Pull off turbines downwards (press support lever).
The frame around the maintenance slot number will disappear from the display.
• Wipe off the instruments using a damp cloth soaked with disinfectant. For Germany we recommend using
disinfectants listed by the VAH (such as SciCan OPTIM BLUE).
• Immediately following this treatment, the instrument must be treated further in accordance with the manufacturer's
instructions on reprocessing (such as disinfection/sterilization in the STATIM).
• Always perform a function test before using the instruments on patients. Some liquid may come out.