8 Technical Data
Page 26
STATMATIC smart 1,2,3 EN
8 Technical Data
Device:
Device dimensions:
Width:
Depth:
Height:
300 mm (11.8 inches)
280 mm (11.0 inches)
345 mm (13.6 inches)
Weight:
7.3 kg (15.8 lbs)
Power ratings:
Rated voltage:
Rated frequency:
Rated power:
100 230 V 10%
50 / 60 Hz
45 VA max.
Protection class:
I
Protection:
covered
Installation category:
Cat. II
Contamination level:
P 2
Ambient temperature:
5 35 °C (41 95 °F)
Noise level:
60 (65) dB(A)
Relative air humidity:
85 % max.
Max. operating altitude:
2000 m above sea level
Operating location:
only indoors
Compressed air:
4.5 6 bar (65 87 psi)
Air consumption:
40 Nl/min
Compressed air quality:
Residual oil content:
Residual dust:
Residual water:
max. 0.1 mg/m³
max. 1 µm; max. 1 mg/m³
max. 0.1 g/m³ at -40 °C (-40 °F)
Power fuses:
Rated current:
Rated voltage:
Triggering characteristic:
Breaking capacity:
2 A
250 V
Large (T)
H
Number of ISO 3964 integrated couplings:
Type 1
Type 2
Type 3
1
2
3
Device: Storage conditions Transport conditions
Ambient temperature:
5 35 °C C (41 95 °F)
-30 50 °C (-22 122 °F)
Relative air humidity:
5 95 %
5 - 95 %
Compressed air:
700 - 1060 hPa
700 - 1060 hPa
Empty the media containers prior to transport! (see Chapter 4 "Maintenance")
Media: Storage conditions Transport conditions
Ambient temperature dental oil:
5 50 °C (41 122 °F)
-30 50 °C (-22 122 °F)
Ambient temperature cleaner:
5 35 °C (41 95 °F)
-30 50 °C (-22 122 °F)
Relative air humidity:
5 - 95 %
5 - 95 %
Compressed air:
800 - 2500 hPa
800 - 2500 hPa