4 Maintenance
Page 16
STATMATIC smart 1,2,3 EN
4 Maintenance
Checking adapter fixation
Recommended:
- After installation
- During operation: 1x week
Check the MULTIflex receptacle (and/or other turbine adapter) for firm seating..
The clamping nut must be tight.
Retighten clamping nut, if required.
Check the output quantity on all cleaning slots using a testing adapter intended
for the coupling.
Please note that only STATMATIC smart testing adapters can be used. STATMATIC
PLUS testing devices are not suitable.
Type: 2
Checking the O-rings
Recommended:
- During operation: 1x week
Check all O-rings on the adapters for damage
Replace damaged O-rings
Functional testing of flow rate using the testing adapter
Recommended:
- Upon initial commissioning
- After changing an instrument coupling
- During operation: 1x day before use
Attach the matching testing adapter to the corresponding instrument couplings
(such as INTRAmatic, DIN ISO EN Coupling).
Close the front cover.
Press the START button.
After completion of the program, check the filling quantity in each of the testing
adapters.
The fluid must be above the mark ("Min.") !
If the filling quantity is below the mark, please consult the SciCan customer
service.
Please note that only STATMATIC smart testing adapters can be used. STATMATIC
PLUS testing devices are not suitable.
Pressure test:
Recommended:
- Upon initial commissioning
- during operation: 1x day, preferably before starting work
Read off the "operating pressure" using the pressure gage.
Pressure should be between 4.5 bar and 4.3 bar!
Type: 3
Changing the adapter
The loosening of the clamping nut allows for the exchangeable adapter to be removed. Insert the desired adapter and
tighten it again using the clamping nut.
Please note that only white STATMATIC smart / STATMATIC PLUS adapters can be used. Adapters (black) of the
STATMATIC maintenance unit are not suitable.
Check the output quantity on the adapter using a testing adapter intended for the coupling.
Please note that only STATMATIC smart testing adapters can be used. STATMATIC PLUS testing devices are not
suitable.