4 Maintenance
Page 20
STATMATIC smart 1,2,3 EN
Warning message: Servicing required soon
This warning message appears if more than 25,000 regular and contra-angle handpieces and turbines have been
cleaned and maintained for with the STATMATIC smart. Please contact the service department. You can clean a further
3,300 regular and contra-angle handpieces and turbines.
Error message: Servicing required
This error message appears if more than 28,300 regular and contra-angle handpieces and turbines have been cleaned
and maintained for with the STATMATIC smart. Please contact the service department.
Transport
• If the device was already filled with dental oil and cleaner, the media must be emptied out fully through the drain hoses.
• Empty the media system (see p. 20)
• Unscrew the safety screw of the media aeration and blow compressed air into the valve several times. Tighten the
screw again.
• Place the accompanying seal plugs in the filling funnel of the media container (see fig.)
• Close the flaps.
• If possible, send in the original box.