20
Command Response Purpose
OZT
OK troy ounces selected
1
Places balance in Troy Ounces.
OZ
OK ounces selected
1
Places balance in Ounces (advp)
LB
OK pounds selected
1
Places balance in Pounds.
KG
OK kilograms selected
1
Places balance in Kilograms.
MG
OK milligrams selected
1
Places balance in milligrams.
A
OK A selected
2
Places balance in special unit of weight.
TAVG
OK Time averaging ON Clears the TBAR registers and enters the
TBAR mode.
RESET_TAVG
OK Time averaging reset Clears the TBAR registers. If the balance is in
the TBAR mode, the process keeps running.
XAVG
#N XX.XXX U
3
Accumulates the current weight in the XBAR
registers.
RESET_XAVG
None Clears the XBAR registers.
GLP
Sends GLP header Puts balance in the GLP mode.
ISO
Sends ISO header and waits
for target weight entry
Puts balance in ISO mode.
TARG XX.XXX
4
XX.XXX g TARGET EN-
TERED
Enter target weight value in grams.
SEND
Sends weight, difference be-
tween weight and target, and
signature line
Terminates ISO mode.
ASON or ASOFF
Output the current weight at
the interval specified by
ASINT
Turns auto send on or off. When auto send is
on, a time interval must be entered. See SETA-
SINT
ASINT n Auto Send Interval n Seconds Sets the auto send interval in Seconds
CSON or CSOFF
Output the current weight at
every 200mS
Turns continuous send on or off.
SSON or SSOFF
Output the current weight
when the OK annunciator ap-
pear in the display.
Turns send when stable on or off.
READCLOCK
Current date and time
RCL_XAVG
Mean XX.XXX U
3
Maximum XX.XXX U
3
Minimum XX.XXX U
3
StdDev XX.XXX U
3
CV XX.XXXX %
N Integer Samples
Sends statistics of samples