20
Command Response Purpose
OZT↵
OK troy ounces selected
1
Places balance in Troy Ounces.
OZ↵
OK ounces selected
1
Places balance in Ounces (advp)
LB↵
OK pounds selected
1
Places balance in Pounds.
KG↵
OK kilograms selected
1
Places balance in Kilograms.
MG↵
OK milligrams selected
1
Places balance in milligrams.
A↵
OK A selected
2
Places balance in special unit of weight.
TAVG↵
OK Time averaging ON Clears the TBAR registers and enters the
TBAR mode.
RESET_TAVG↵
OK Time averaging reset Clears the TBAR registers. If the balance is in
the TBAR mode, the process keeps running.
XAVG↵
#N XX.XXX U
3
Accumulates the current weight in the XBAR
registers.
RESET_XAVG↵
None Clears the XBAR registers.
GLP↵
Sends GLP header Puts balance in the GLP mode.
ISO↵
Sends ISO header and waits
for target weight entry
Puts balance in ISO mode.
TARG XX.XXX↵
4
XX.XXX g TARGET EN-
TERED
Enter target weight value in grams.
SEND↵
Sends weight, difference be-
tween weight and target, and
signature line
Terminates ISO mode.
ASON↵ or ASOFF↵
Output the current weight at
the interval specified by
ASINT
Turns auto send on or off. When auto send is
on, a time interval must be entered. See SETA-
SINT
ASINT n Auto Send Interval n Seconds Sets the auto send interval in Seconds
CSON↵ or CSOFF↵
Output the current weight at
every 200mS
Turns continuous send on or off.
SSON↵ or SSOFF↵
Output the current weight
when the OK annunciator ap-
pear in the display.
Turns send when stable on or off.
READCLOCK↵
Current date and time
RCL_XAVG↵
Mean XX.XXX U
3
Maximum XX.XXX U
3
Minimum XX.XXX U
3
StdDev XX.XXX U
3
CV XX.XXXX %
N Integer Samples
Sends statistics of samples