EasyManua.ls Logo

ScienTECH SA Analytical SA 210 - Page 20

Default Icon
32 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
20
Command Response Purpose
OZT↵
OK troy ounces selected
1
Places balance in Troy Ounces.
OZ↵
OK ounces selected
1
Places balance in Ounces (advp)
LB↵
OK pounds selected
1
Places balance in Pounds.
KG↵
OK kilograms selected
1
Places balance in Kilograms.
MG↵
OK milligrams selected
1
Places balance in milligrams.
A↵
OK A selected
2
Places balance in special unit of weight.
TAVG↵
OK Time averaging ON Clears the TBAR registers and enters the
TBAR mode.
RESET_TAVG↵
OK Time averaging reset Clears the TBAR registers. If the balance is in
the TBAR mode, the process keeps running.
XAVG↵
#N XX.XXX U
3
Accumulates the current weight in the XBAR
registers.
RESET_XAVG↵
None Clears the XBAR registers.
GLP↵
Sends GLP header Puts balance in the GLP mode.
ISO↵
Sends ISO header and waits
for target weight entry
Puts balance in ISO mode.
TARG XX.XXX↵
4
XX.XXX g TARGET EN-
TERED
Enter target weight value in grams.
SEND↵
Sends weight, difference be-
tween weight and target, and
signature line
Terminates ISO mode.
ASON↵ or ASOFF↵
Output the current weight at
the interval specified by
ASINT
Turns auto send on or off. When auto send is
on, a time interval must be entered. See SETA-
SINT
ASINT n Auto Send Interval n Seconds Sets the auto send interval in Seconds
CSON↵ or CSOFF↵
Output the current weight at
every 200mS
Turns continuous send on or off.
SSON↵ or SSOFF↵
Output the current weight
when the OK annunciator ap-
pear in the display.
Turns send when stable on or off.
READCLOCK↵
Current date and time
RCL_XAVG↵
Mean XX.XXX U
3
Maximum XX.XXX U
3
Minimum XX.XXX U
3
StdDev XX.XXX U
3
CV XX.XXXX %
N Integer Samples
Sends statistics of samples