EasyManua.ls Logo

Scientifica IVM - Page 20

Scientifica IVM
44 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
2| 4.0 Operation Guide
Make sure that all the cables are loose throughout the range
of motion and stages.
Is the drift detected with the perfusion running?
(test for longer than 10 minutes)
Perfusion solution with a temperature different to the room
can lead to thermal expansion of the pipette – leave some
time for the pipette to adjust to new temperatures.
Fast perfusion may cause drift of the pipette.
Ensure the vacuum source is strong enough to break the
surface tension of the fluid in the bath
Is the microscope firmly mounted to the table top?
Many microscopes have rubber feet underneath. These may
degenerate with time and cause the microscope to be
unstable. Ideally the microscope should be bolted down onto
the table top.
Is the cooling headstage switched off?
Some amplifiers have a cooling facility on the headstage
(which is normally located on the amplifier front or back
panel). This may lead to expansion of the headstage and
subsequent drift.
You should allow half an hour after switching off before
re-testing stability
Recording chamber: are there any cables, wires or tubing
connected to the chamber?
Such cables could implement an external force causing
movement of the chamber.
If using brain slices, ensure that the harp use to hold
down the slice does so efficiently.
Check for air currents around the system
Open windows or air conditioning can give rise to air
currents.
Cover the system with a sheet or plastic bag to eliminate
any air currents then test for drift.
If you are still experiencing drift difficulties after following these
steps, please contact us for further assistance.

Table of Contents

Related product manuals