5.1 Troubleshooting
5.1.1 No movement
Check Scientifica logo is illuminated on the front of the rack
Check all cables and connectors for damage or loose
connections
5.1.2 Drift
Allow the system to warm up
With most electrical components on the rig needing a warm-up
period, it is good practice to allow a certain warming-up time
after switching on the power of the set-up (up to one hour).
Check all bolts are tightened
Is the Headstage or probe holder attached correctly?
Make sure that the headstage is firmly attached
Be careful that you don’t over tighten the clamp, because this
may cause expansion and may also be a source of drift.
Check that the headstage-mounted sliding bracket is
clamped at the end of its travel.
Is the pipette holder attached firmly and the O-ring new and of
appropriate size?
With pipettes being exchanged very often, there is a lot of
wear on the O-ring inside the pipette holder which may lead
to it not holding the pipette firmly.
Also make sure that there is no tension on the pipette suction
tube
Ensure that the AgCl wire inside the pipette is not touching
the front end of the electrode.
Are the 3 linear stage wires and the rotation sensor wires
supported?
There should be a soft loop on all wires from the manipulator,
ideally fixed to a point close to the manipulator mount.
The cables can also be fixed to the Faraday cage above the
manipulator, but only if the Faraday cage is isolated from the
floor.