Indication Warning description
Warning: Review
Not selecting in review data list.
Solution
Press ‘Send selected result’ after selecting the
result.
Warning: Unregistered Operator ID
Entered operator ID does not be registered.
Solution
Add operator ID in set mode.
Warning: Out of Paper
There is no printer paper in analyzer.
Solution
Rell printer paper.
Warning: Incorrect IP address
Check IP address.
Solution
Conrm that entered IP address and input
correct IP address.
Warning: Full memory
Store the maximum of 3000 data.
Solution
If the next test result is measured after this
warning, the oldest result is erased.