Error Messages
Indication Error description
E01: Test Device Error
The test device is damaged or inserted
improperly.
Solution
Discard the test device and re-test with a new
test device and a new specimen.
E02: Blood Specimen Error
An insufcient blood has been applied.
Solution
Re-test with the new test device with an
enough specimen, ensuring that blood is
placed to the narrow channel in the top edge
of the test device.
E03: Expired Test Device
The test devices are expired.
Solution
Re-test with the new test device with non-
expired date.