6. When the countdown reaches 0, your result appears on the
screen. If using a lipid pro le test strip, use the left and right
arrows to view the results for the di erent parameters.
BL
7. When the test is complete, pull out the used test strip and
discard in a biohazardous waste container. The Analyzer shuts o
automatically 3 seconds after you remove the test strip.
Do not open the Measurment chamber  ap during the test.
If you do not close the measurement chamber  ap immediately after applying blood, the analyzer display  ashes the close
ap icon with countdown and beep.
If you do not close the  ap within 20 seconds of applying the blood, the Analyzer displays the ' E-7 ' message.
The blood drop must  ll the entire blood application hole.
Example of not
enough blood
Example of
enough blood
Example of not
enough blood