EasyManua.ls Logo

SD Biosensor STANDARD LipidoCare - Page 19

SD Biosensor STANDARD LipidoCare
45 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
19
6. When the countdown reaches 0, your result appears on the
screen. If using a lipid pro le test strip, use the left and right
arrows to view the results for the di erent parameters.
BL
7. When the test is complete, pull out the used test strip and
discard in a biohazardous waste container. The Analyzer shuts o
automatically 3 seconds after you remove the test strip.
Do not open the Measurment chamber  ap during the test.
If you do not close the measurement chamber  ap immediately after applying blood, the analyzer display  ashes the close
ap icon with countdown and beep.
If you do not close the  ap within 20 seconds of applying the blood, the Analyzer displays the ' E-7 ' message.
The blood drop must  ll the entire blood application hole.
Example of not
enough blood
Example of
enough blood
Example of not
enough blood

Table of Contents