41
2. Test strip for Lipid
Measuring Range
TC: 100 - 450 mg/dL ( 2.59 - 11.64 mmol/L )
TG: 45 - 650mg/dL ( 0.51 - 7.34 mmol/L )
HDL: 25 - 95 mg/dL ( 0.65 - 2.46 mmol/L )
Sample
Fresh capillary whole blood or
venous whole blood, serum or plasma
Sample Size 10 µl(Single) / 35 µl (Lipid Prole)
Test Time 3 minutes
Test Strip Storage Temperature 2 - 32°C (36 - 90°F)
Hematocrit
TC and TG parameter: 30 - 50%
HDL parameter: 30-52%
Testing Temperature 18 - 32°C (64 - 90°F)
3. Test strip for Glucose
Measuring Range 10 - 600 mg/dL (0.6 - 33.3 mmol/L)
Sample Fresh capillary whole blood
Sample Size 0.9 µl
Test Time 5 seconds
Test Strip Storage Temperature 2 - 32°C (36 - 90°F)
Hematocrit 20 - 60%
Testing Temperature 10 - 45°C (50 - 113°F)
Altitude Up to 3,776m(12,388 ft)
Calibration Plasma-equivalent
4. Electromagnetic Compatibility
This Analyzer meets the electromagnetic immunity requirements as per EN ISO 15197 Annex A. The chosen basis for electrostatic discharge immunity testing
was basic standard IEC 61000-4-2. In addition, it meets the electro -magnetic emissions requirements as per EN 61326. Its electromagnetic emission is thus
low. Interference from other electrically driven equipment is not to be anticipated.