15
symbol. This number and parameter name must match the codechip. If it does not, repeat the steps above (1-2).
BL
BL
BL
Lipid Profile TC TG
A codechip is included in each test strip package. Store the lipid test strips together with the codechip in the package.
Glucose Test Strip
A codechip is not necessary for glucose testing.
2. Preparing for measurement
Lipid Test System - components required:
• STANDARD LipidoCare Analyzer
• Lipid test strips for the desired parameter with the related codechip
• Lancets and Lancing device
• Tool for collecting the blood such as the STANDARD Ezi tube+ or micropipette
Only use STANDARD LipidoCare Lipid Test Strip and either STANDARD LipidoCare Blood Glucose Test Strip or SD Codefree Blood Glucose test
strip.
Precaution
Do
• Read the test strip package insert.
• Operate the Analyzer at the acceptable test-specic temperatures.
- For Lipid testing 18-32 °C (64-90°F)
- For Glucose testing 10-45 °C (50-113°F)
• Place the Analyzer on a level surface or hold it steady in your hand.
• Keep the guidelines for cleaning.
• Make sure that all display elements are displayed.
• Keep the STANDARD Lipid Check Strip away from sunlight.
Do not
• Use lipid test strips beyond the expiration date, this may cause inaccurate results.
• Touch or remove the test strip during actual measurement.
• Subject the Analyzer to sudden movements during a measurement.