2. Insert the STANDARD Lipid Check Strip.
3. The check result will appear on the screen in just 10 seconds. If there is no problem with the Analyzer, 'OK' will appear on the screen; if there is a
problem with the analyzer, ‘EEE’ will appear on the screen and you should contact your local distributor.
2. How to Use the STANDARD Glucose Check Strip
1. Insert the STANDARD Glucose Check Strip (with the words check strip’ facing upwards) into the glucose test strip slot. The Analyzer will turn on
automatically.
2. The check result will appear on the screen in 5 seconds. If there is no problem with the Analyzer, 'OK' will appear on the screen; if there is a problem
with the analyzer, ‘EEE’ will appear on the screen and you should contact your local distributor.
BL
CHAPTER 9. Maintenance and Troubleshooting
1. Cleaning your Analyzer
Caring for the STANDARD LipidoCare Analyzer is easy. If you need to clean it, follow these guidelines carefully to help you get the best performance possible;
To prevent malfunction of the Analyzer, keep the test strip slot free of blood, moisture, dirt, or dust. Use a lint-free cloth dampened with water to clean the
Analyzer. Ensure the cloth is damp, and not wet. Do not use an abrasive cloth or antiseptic solution, as these may damage the display screen.
A clean optical measuring system is a basic prerequisite for determining precise measured values.
Therefore clean the Analyzer regularly and immediately after it becomes dirty. Always switch o the Analyzer before cleaning it.
Use only the following items for cleaning:
Ordinary lint-free cotton buds
Ordinary lint-free tissues
Ordinary disinfecting tissues
Do not use any disinfectant sprays or tissues or cotton buds which are dripping wet as the liquid may enter the Analyzer and damage it.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the SD Biosensor STANDARD LipidoCare and is the answer not in the manual?

SD Biosensor STANDARD LipidoCare Specifications

General IconGeneral
BrandSD Biosensor
ModelSTANDARD LipidoCare
CategoryMedical Equipment
LanguageEnglish