4. Running a test - Glucose
Test Procedure
1. Remove a new glucose test strip from its container. Be sure to tightly
replace the container cap immediately after removing a glucose test
strip.
2. Insert the glucose test strip into the 'Glucose Test Strip Slot' of the
Analyzer. The Analyzer will automatically enter the glucose testing
mode.
BL
3. When the blood drop symbol  ashes, you are ready to perform a
glucose test.
BL
4. Obtain a blood sample using the lancet and lancing device.
0.9µL
actual size
5. Hold your  nger to the tip of the glucose strip until the yellow window is completely  lled with blood. Do not place the blood drop on top of the
glucose strip.
6. The blood will be drawn into the glucose strip automatically. If the beeper is turned on, the Analyzer will beep to let you know the glucose test is beginning.
7. When blood is applied to the glucose strip, the display counts down from 5 to 1 and your result appears on the display in just 5 seconds.
8. When the test is done, pull out the used glucose test strip. The Analyzer shuts o automatically 3 seconds after you remove the test strip.
Narrow Channel
STEP 1
STEP 2
Gently touch the glucose test strip to
the blood drop. Do not press the
glucose test strip against the finger.
Proper glucose test
strip appearance.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the SD Biosensor STANDARD LipidoCare and is the answer not in the manual?

SD Biosensor STANDARD LipidoCare Specifications

General IconGeneral
BrandSD Biosensor
ModelSTANDARD LipidoCare
CategoryMedical Equipment
LanguageEnglish