37
Indication Warning description
BL
Warning
Your test result is lower than the measuring range.
*This indication is for Glucose.
Indication Warning description
BL
Warning: Low Battery
At this time, battery is getting low but you can still perform about 50 tests.
Indication Warning description
BL
Warning: Replace Battery
Battery power is low.
If the battery is empty and you press the power button, the battery icon will ash for 10 seconds and then the
analyzer will turn o automatically.
2. Error message
Indication Error description
In case of lipid test strip
In case of glucose test strip
Error: Strip Error
The lipid or glucose test strip is dirty, used, damaged or inserted improperly.
Solution
Discard this test strip and test again using a new test strip.