38
Indication Error description
In case of lipid test strip
In case of glucose test strip
Error: blood sample Error
An insucient amount of blood was applied.
Solution
Discard this test strip and test again using a new test strip with a larger sample.
Indication
Error
description
Error: Expired strip (Only for a Lipid test strip)
The Lipid test strips are expired.
Solution
Discard this test strip and test again using a new lipid test strip that has not expired. If the strips are not expired
and you still get E-3 then this means the year and/or date are not set correctly. Please refer to the section on
setting the Analyzer.
Indication Error description
Error: Temperature Error
If the environmental temperature is above or below the operation range of the Analyzer, a thermo Analyzer icon
will appear on the display.
Solution
Move to an area between 18-32°C(64-90°F) for lipid or 10-45°C (50-113°F) for glucose, wait for 30 minutes, and
perform a test. Do not articially heat or cool the Analyzer.
Indication Error description
Error: Communication Error
The communication between Analyzer and computer or printer is failed.
Solution
Connect the Analyzer and external device again.
If the error message persists, please contact your local distributor.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the SD Biosensor STANDARD LipidoCare and is the answer not in the manual?

SD Biosensor STANDARD LipidoCare Specifications

General IconGeneral
BrandSD Biosensor
ModelSTANDARD LipidoCare
CategoryMedical Equipment
LanguageEnglish