39
Indication Error description
Error: Code mismatching Error
The inserted test strip and codechip do not match.
Solution
Change the appropriate test strip or codechip.
Indication Error description
Error: Flap Open
The chamber ap of the Analyzer is open during measurement.
Solution
Close the chamber ap and do not touch or open the ap during measurement.
Indication Error description
Error: Blood Error
Measurement chamber ap is closed for testing without blood sample.
Solution
Re-test with sucient blood sample.
Indication Error description
Error: Communication Error with codechip
The Analyzer cannot communicate with the inserted codechip.
Solution
Re-insert the codechip. If the error message persists, please contact your local distributor.
Indication Error description
Error: Internal Error
There is an internal error of the analyzer.
Solution
Turn the Analyzer o and then back on again. If the error message persists, please contact your local distributor.