Weighing with pre-Tare
If the pre-Tare function is activated, a stored weight value is deducted from the currently
measured weight. This function is useful, for example, if the weight of a wheelchair is not
to be displayed.
Three different tare weights (wheelchair weights) can be stored.
Press the Start key.
se(a, 8.8.8.8.8. and 0.00. appear
consecutively in the display. After that,
the scale is automatically set to zero
and ready for operation.
Press the FUNC key.
The last function used appears in the
display.
An arrow on the left-hand side of the
display indicates which memory is set.
Use the arrow keys to select the de-
sired memory for storing the additional
weight or where a weight may already
be stored.
Then press FUNC. In the display you
will see the tare value from the memory.
You can use the arrow keys to adjust
the value in increments according to
the graduations displayed.
Once the correct value is set, press the
FUNC key again.
The pre-Tare function is now activated.
Move onto the scale in the wheelchair
associated with that memory. Read the
weighing result off the digital display.
Press an arrow key if you would like to
see the current weight value briefly.
To return to normal weighing mode,
press the FUNC key.
To switch off the scale, press the Start
key again.
Tip: note weight and the memory in
which the weight is stored on your
wheelchairs.