Maintenance seca
14.11.12/RLEU/ARI/MRE 15/40 30-34-00-812c
less than 1.5 seconds). change the value on the display due to
the low display resolution. Only the
rounded value is output.
Repeat this step until the measured
weight matches the test weight (format:
xxx.0 kg).
The display indicates the test weight
(format: xxx.0 kg).
To correct this rounding error, decrement
or increment the value five times by 10 g.
The display indicates the exact test
weight (format: xxx.0 kg).
Press the tare/hold key until the display
shows “Sto”.
The display reads “Sto” and the scale
switches off automatically.