2.1.9 Overview of Adjusting a Baby Scale with 10g Graduations
[374|376|378]
Action carried out by the user
Result
The scale is switched off. No values or characters on the display.
Press the tare/hold key and keep it
pressed while simultaneously pressing
the start key.
The contents of the calibration counter
flash on the display for 18 seconds.
Release both keys. The contents of the calibration counter
flash on the display for 18 seconds.
Press the tare/hold key for more than 1.5
seconds.
The system switches to adjustment
selection mode and “CAL” flashes on the
display.
18 seconds have passed and no key has
been pressed.
The scale switches off.
Press the tare/hold key for more than 1.5
seconds.
"dec" appears on the display (flashing)
and then the current weight (0.000 - also
flashing).
Place the test weight on the scale. The measured weight of the test weight
appears on the display (flashing).
Decide whether to decrement or to
increment:
Incrementing:
Press the tare/hold key for more than 1.5
seconds.
Decrementing: No action required as the
scale already is in dec mode.
Incrementing mode is activated; "inc"
appears on the display.
No output on the display.
Decrementing or incrementing: The measured weight is decremented or