14.11.12/RLEU/ARI/MRE 24/40 30-34-00-812c
3.4 Scale Error Messages
[374|376|378|703|704|284|285|634|635|644|645|656|657|664|665|
674|675|676|677|684|685|954|957|959|963]
Display: Er:x:nr Electronics detecting a fault.
X : Module number
No Fault number
Scale displays Cause Remedy
seca serva
diagnosis: replace
load cell or
electronics as
required
Er:X:11
Load cell defective,
supply lines damaged or
electronics defective
Replace the
electronics
Er:X:12 Initial load too high
Check the initial
load
Er:X:15
Readjustment error
Reference weight
outside the
permitted range or
operating error
Er:X:16 Error when determining
the switch-on zero point
Avoid subjecting
the scale to
vibrations during
start-up
Er:X:30
Data transfer interrupted Check the bus
connection/
modular cable (if
provided)