94 •
Only the "SYS" item is
visible in the "rF" menu.
The wireless module is deactivated.
- Activate wireless module (see "Activating the
wireless module (SYS)" on page 87).
Only the "SYS" and "Lrn"
items are visible in the
"rF" menu.
The wireless module is activated and no
wireless group is set up.
- Set up wireless group (see "Setting up a
wireless group (Lrn)" on page 87).
The items "APrt" and
"Time" are not visible in
the "rF" menu.
No wireless printer registered to the wireless
group.
- Register the wireless printer to the wireless
group via the "Lrn" menu item (see "Setting
up a wireless group (Lrn)" on page 87).
After calling up the
menu, the "rF" item is
not displayed.
The scale’s wireless module is defective.
- Inform seca service.
The display "Er:X:11"
appears.
The scale has too high a load or too high a load in
one corner.
- Take the load off the scale or distribute the
weight more evenly.
- Restart the scale.
The display "Er:X:12"
appears.
The scale has been switched on with too high a
load.
- Remove the load from the scale.
- Restart the scale.
The display "Er:X:16"
appears.
The scale was caused to oscillate, the zero point
could not be determined.
- Restart the scale.
The Enter key
(send/print) is pressed
and the display "Er:X:71"
appears.
No data transmission possible, wireless module
deactivated.
- Activate wireless module (see "Activating the
wireless module (SYS)" on page 87).
The Enter key
(send/print) is pressed
and the display
"Er:X:72" appears.
No data transmission possible, no wireless group
set up.
- Set up wireless group (see "Setting up a
wireless group (Lrn)" on page 87).
Problem Cause / Solution

Table of Contents