Technical data • 97
11.2 Technical data, weighing
Medical device in accordance with Direc-
tive 93/42/EEC
Class I with measuring function
EN 60601-1:
• insulated device, protection class II
• electrical medical device, type B
Wireless transmission
•Frequency band
• Transmission power
• Standards applied
2.433 GHz - 2.480 GHz
< 10 mW
EN 300 328
EN 301 489-1
EN 301 489-17
General technical data seca 665
Technical data, weighing seca 665
Calibration in accordance with Directive
2014/31/EU
Class III
Maximum load
• Partial weighing range 1
• Partial weighing range 2
200 kg
300 kg
Minimum load
• Partial weighing range 1
• Partial weighing range 2
2.0 kg
4.0 kg
Increments
• Partial weighing range 1
• Partial weighing range 2
100 g
200 g
Tare range 300 kg
Accuracy on initial calibration
• Partial weighing range 1, to 50 kg
• Partial weighing range 1, 50 kg to 200 kg
• Partial weighing range 2, to 100 kg
• Partial weighing range 2, 100 kg to 300 kg
±50 g
±100 g
±100 g
±200 g