74 •
9. FUNCTION CHECK
Perform a function check prior to each use.
A complete function check includes:
visual inspection for mechanical damage
checking the alignment of the device
visual and function check of the display elements
function check of all the controls shown in the
section entitled “Overview”
function check of optional accessories
If you notice any faults or deviations during the function
check, first try to resolve the error with the aid of the
section entitled “What do I do if ...” in this document.
CAUTION!
Personal injury
If you notice any faults or deviations during the
function check which cannot be resolved with
the aid of the section entitled “What do I do if
...” in this document, you may not use the
device.
Have the device repaired by seca service or
by an authorized service partner.
Follow the section entitled “Servicing” in this
document.
10.WHAT DO I DO IF …?
Fault Cause/Remedy
… no weight is dis-
played when scale is
loaded?
The scale is not connected to the power supply.
- Check if the scale is switched on.
- Check whether the battery block is inserted
and that the plug contacts are properly con-
nected.
… 0.000 does not appear
before weighing?
A load was placed on the scale before it was
switched on.
- Remove load.
- Switch scale off and back on again.
… a segment lights up
continuously or not at
all?
The relevant segment is faulty.
- Please notify maintenance service.
is displayed?
Battery block power is running down.
- Charge battery block.

Table of Contents

Other manuals for Seca 727

Related product manuals