EasyManua.ls Logo

Seca 727 - 11 Servicing

Seca 727
293 pages
Go to English
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
76 •
11.SERVICING
On leaving the factory, your seca scale has an accu-
racy of ±0,15 % or better. To preserve this level of
accuracy, the product must be set up with care and
serviced regularly. Depending on how frequently the
scale is used, we recommended servicing at intervals
of 3 to 5 years.
NOTICE!
Incorrect measurements as a result of poor
servicing
Have servicing and repairs carried out
exclusively by seca service or by an
authorized service partner.
You can find service partners in your area at
www.seca.com or by sending an e-mail to
service@seca.com.
… the display “Er:X:11”
appears?
The scale has too high a load, or too great a load
is applied in one corner.
- Reduce load on scale or distribute weight
more evenly.
- Restart scale.
… the display “Er:X:12”
appears?
The scale has been switched on with too great a
load.
- Remove load.
- Restart scale.
… the display “Er:X:16”
appears?
Oscillations have been induced in the scale, pre-
venting determination of the zero point.
- Do not touch the trolley or table supporting
the scale.
- Restart scale.
… when the Enter key
(send/print) is pressed,
the “Er:x:71” display
appears?
Data transmission not possible, wireless module is
disabled.
- Activate wireless module (“Activating the wire-
less module (SYS)” on page 68).
… when the Enter key
(send/print) is pressed,
the “Er:x:72” display
appears?
Data transmission is not possible, no wireless
group set up.
- Set up wireless group (“Setting up a wireless
group (Lrn)” on page 69).
Fault Cause/Remedy