Technical data • 77
12.TECHNICAL DATA
12.1 General technical data
General technical data seca 727
Dimensions
•Depth
•Width
•Height
320 mm
551 mm
151 mm
Weight approx. 7.1 kg
Ambient conditions, operation
• Temperature
• Air pressure
• Humidity
+10 °C to +40 °C (50 °F to 104 °F)
700 - 1060 hPa
30 % - 80 %, no condensation
Ambient conditions, storage
• Temperature
• Air pressure
• Humidity
-10 °C to +65 °C (14 °F to 149 °F)
700 - 1060 hPa
0 % - 95 %, no condensation
Ambient conditions, transport
• Temperature
• Air pressure
• Humidity
-10 °C to +65 °C (14 °F to 149 °F)
700 - 1060 hPa
0 % - 95 %, no condensation
Height of digits 11 mm
Power supply Battery block, power pack
Power consumption
• with deactivated wireless module and
without background lighting
•with activated wireless module and per-
manent background lighting (100 %)
approx. 47 mA
approx. 100 mA
Maximum runtime in battery mode
• with deactivated wireless module and
without background lighting
• with wireless module activated
approx. 2880 minutes
Power pack recommended
Medical device in accordance with
Directive 93/42/EEC
Class I with measuring function
EN 60 601-1:
• Class II insulated appliance:
• Type B electrical medical device:
Type of protection IP20
Duty cycle Continuous duty